3166 & 3167 TEM Horn Antennas

The ETS-Lindgren Model 3166 and 3167 TEM Horn Antennas generate homogeneous fields for immunity testing per the IEC/EN 61000-4-39 standard.
The ETS-Lindgren Model 3166 and 3167 TEM Horn Antennas generate homogeneous fields in the frequency range of 380 MHz to 6 GHz specifically for immunity testing in close proximity as per the IEC/EN 61000-4-39 standard. The 3166 model operates from 380 MHz to 3 GHz with less than 7: 1 VSWR from 400 MHz to 700 MHz and 3: 1 VSWR from 700 MHz and above. The 3167 operates from 3 GHz to 6 GHz with less than 2: 1 VSWR.
Key Features
- Generates homogeneous fields for immunity testing
- Close proximity immunity testing
- IEC/EN 61000-4-39 standard
Product Details
Physical Specifications
3166 Physical Specifications
- Width: 21.6 cm (8.5 in)
- Depth: 47.6 cm (18.75 in)
- Height: 13.2 cm (5.2 in)
- Approximate Weight: 3.6 kg (8 lb)
3167 Physical Specifications
- Width: 5.7 cm (2.24 in)
- Depth: 17.3 cm (6.8 in)
- Height: 3.2 cm (1.26 in)
- Approximate Weight: 0.2 kg (0.4 lb)
Electrical Specifications
3166 Electrical Specifications
- Application: IEC/EN 6100000-4-39
- Frequency Range: 380 Mhz to 3 GHz
- VSWR: 7:1 (400 MHz to 700 MHz), 3:1 700MHz to 3 GHz)
- Maximum Continuous Power: 250 W
- Impedance (Nomincal): 50
- Connector: N Type, Female
3167 Electrical Specifications
- Application: IEC/EN 6100000-4-39
- Frequency Range: 3 GHz to 6 Ghz
- VSWR: 2:1
- Maximum Continuous Power: 250 W
- Impedance (Nomincal): 50
- Connector: N Type, Female